Tools, Technologies and Training for Healthcare Laboratories

The 2021 African QC Survey Results

The Global QC Survey of 2021 collected responses Africa, revealing significant differences in the laboratories of the developing world.

 The 2021 African QC Survey Results

Sten Westgard, MS
December 2021

 

 [This survey was completed with the support and partnership of Technopath Clinical Diagnostics.]

The demographics

 2021 QC survey Africa q4 lab type

Public hospitals dominated our African respondents.

2021 QC survey Africa q5 lab volume

Smaller labs dominated the responses of this region. Almost no very large and mega-labs.

 2021 QC survey Africa q6 lab regulations

ISO 15189 and other local regulations are the strongest determining factor, while CLIA and CAP has very, very limited impact.

 2021 QC survey Africa q7 lab testing choice

 The respondents focused most on the QC performed in their biochemistry testing.

Now, how do these labs practice QC?

The QC Set Up

 2021 QC survey Africa q8 2sds

Two-thirds of labs are using 2 SD limits in Africa. This is higher than the global average.

 2021 QC survey Africa q9 westgard rules

Over 90% of the laboratories from this region use "Westgard Rules" on some or all of their tests. There are no labs that don't use Westgard Rules somewhere. This is a higher rate than the US.

2021 QC survey Africa q10 chart setup 

Nearly two-thirds of laboratories in Africa establish their own mean and SD. Nearly three-quarters use manufacturer ranges, which is a global high. All other practices are very low.

2021 QC survey Africa q11 chart regulations 

Africa has a huge reliance on ISO 15189, the global standard for quality. Very little else influences their control charts.

 2021 QC survey Africa q12 control types

The use of manufacturer controls are extremely high. Independent controls, which represent an additional expense, are used much lower than seen elsewhere in the world. Assayed controls are surprisingly popular. It's not surprising to see lyophilized controls more popular than liquid controls, but not by much. 

2021 QC survey Africa q13 patient data 

Nearly two-thirds of laboratories use no patient data techniques at all. However, given the difficulty in implementing these techniques, it's impressive to see that over 30% of labs use some amount of PBRTQC. There is 13% more use of these patient data techniques in Africa than in the US!

The Real Practice of Running Controls

 2021 QC survey Africa q15 qc frequency

 Once-a-day QC is practices by almost half the labs in Africa, with a surprising high rate of running QC twice a day. The rate of African labs running twice a day is twice the rate of the US.

 2021 QC survey Africa q16 troubleshooting

Laboratories in this region troubleshoot and seek root cause (as well as recalibrate) at the same rate as we see in the US. However, they repeat the control at half the rate of the US! And they run new controls significantly less often than the US. They do, however, recalibrate at significantly higher rates.

 2021 QC survey Africa q17 ooc frequency

Labs in Africa report daily errors at one-sixth the rate of the US. On a given week, labs in Africa are reporting 20% fewer errors than in the US. Does this mean labs in Africa are missing errors, or labs in the US are finding errors that don't exist? 

 2021 QC survey Africa q18 retesting

The retesting rate of all patients is very similar to the US. However, labs in Africa are twice as likely to test selected group of patients (abnormals) than in the US.


 2021 QC survey Africa q19 override

Labs in Africa are four times more likely to override their QC on a regular basis than the US. That's still less than what we've seen in Latin and South America.

The Final Overview

 2021 QC survey Africa q20 manage costs

Labs in Africa have reduced their QC frequency at nearly three times the rate in the US. They are almost three times more likely to have switched to cheaper controls. But two-thirds have done nothing to their QC practices that manages their costs.

Overall, there are lot of practices in this region that are better than what we see in the US. It is a pleasing surprise to see that there are a strong set of optimal QC practices in place in this region.

 

 

 

Let us know what you're interested in!
Please use this form to request more infromation about 

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input