Stamford Hospital Sigma Verification of Performance
Stamford Hospital Laboratory is the first laboratory in Connecticut to reach Sigma Verification.
Stamford Hospital Laboratory Verification of Performance
Verification: January 2, 2019
Duration of Verification: January 2, 2019 through January 31st, 2020
With Stacey DeBlasio, Laboratory Operations Manager
Introduction to Stamford Hospital (from their website):
"Stamford Hospital, residing on the Bennett Medical Center campus, is a 305-bed, not-for-profit hospital providing area residents (Fairfield and Westchester counties) with access to the latest technology using a compassionate, patient-centered care approach.
"Our areas of expertise include: cancer care, heart services, orthopedics and women and children's health. Stamford Hospital is a major teaching affiliate of the Columbia University College of Physicians & Surgeons, with nearly every one of their residents completing rotations at Stamford Hospital."
The laboratory staff of Stamford Hospital
Assays Verified at Stamford Hospital Laboratory
Biochemistry Assay |
C8000 |
C16000 |
Albumin |
Five Sigma |
|
Alk Phos |
Six Sigma |
Six Sigma |
Amylase |
Six Sigma |
Six Sigma |
ALT |
|
Six Sigma |
AST |
|
Six Sigma |
Bilirubin, Direct |
Six Sigma |
Six Sigma |
Bilirubin, Total |
||
C3 complement |
Five Sigma |
|
C4 complement |
Six Sigma |
|
Calcium |
Five Sigma |
Five Sigma |
Chloride |
Five Sigma |
Four Sigma |
Cholesterol |
Six Sigma |
Six Sigma |
CO2 |
Four Sigma |
Four Sigma |
Creatinine |
Six Sigma |
Six Sigma |
Creatinine Kinase (CK) |
|
Six Sigma |
GGT |
|
Six Sigma |
Glucose |
|
|
HDL |
Six Sigma |
Six Sigma |
Haptoglobin |
Five Sigma |
Six Sigma |
IgA |
Four Sigma |
Four Sigma |
IgG |
Six Sigma |
Six Sigma |
Iron |
Six Sigma |
|
Lactate |
Six Sigma |
Six Sigma |
LDH |
Six Sigma |
Six Sigma |
LDL |
|
Six Sigma |
Lipase |
Five Sigma |
Six Sigma |
Lithium |
|
Five Sigma |
Magnesium |
Six Sigma |
Six Sigma |
Phosphorous |
Five Sigma |
Six Sigma |
Potassium |
Six Sigma |
Six Sigma |
Prealbumin |
Six Sigma |
|
Protein, Total |
Five Sigma |
Six Sigma |
Transferrin |
Six Sigma |
|
Triglycerides |
Six Sigma |
Six Sigma |
Urea |
Four Sigma |
Four Sigma |
Uric Acid |
|
Six Sigma |
Immunoassay performance
Immunoassay |
I2000 129829 |
I2000 129833 |
AFP |
Four Sigma |
|
Carbemazepine |
Four Sigma |
|
Folate |
|
Four Sigma |
FSH |
|
Four Sigma |
LH |
|
Six Sigma |
Phenobarbital |
Four Sigma |
|
Prolactin |
|
Five Sigma |
PSA |
Five Sigma |
|
Theophylline |
Four Sigma |
|
Graphic Display of Stamford Hospital Laboratory Sigma Performance
Congratulations to Stamford Hospital Laboratory.
With Peter Zezima, Certified Quality Manager