Tools, Technologies and Training for Healthcare Laboratories

Sigma Verification of Saigon Hoan My Performance

The laboratory of the Hoan My Sai Gon hospital in Ho Chi Minh City, Vietnam has joined the Sigma VP program.

Sigma Verification of the Laboratory of the Hoan My Sai Gon Hospital

Original Verification: December 12th, 2017
Duration of Verification: November 1st, 2020 through December 31st, 2021

"Hoan My Sai Gon (HMSG) hospital is the largest site of Hoan My Medical system which is recognized as the biggest private hospital chain in Vietnam. Hoanmy system includes 11 hospitals and 4 clinics and continue to enlarge the network with many M&A and expansion projects. The laboratory of HMSG hospital plays a very important role as it is the biggest LAB with leading quality positioning and commitment within the network. HMSG LAB was accredited for ISO 15189 certification in May, 2017. Everyday, this LAB serves around 1000 patients with more than 10,000 test of clinical chemistry, immunology, hematology, microbiology and molecular biology. HMSG LAB always keep improving the quality in order to serve their customers better."

Chemistry Assay

Verified on c803866

Verified on c804581

Albumin

-

Six Sigma

ALT

Five Sigma

Six Sigma

Amylase

Five Sigma

Six Sigma

AST

Six Sigma

Six Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Bilirubin, Total

Six Sigma

Four Sigma

C-Reactive Protein

Six Sigma

Six Sigma

Calcium

Six Sigma

-

Chloride

Five Sigma

Six Sigma

Cholesterol

Six Sigma

Six Sigma

Creatinine

Six Sigma

Six Sigma

GGT

Six Sigma

Six Sigma

Glucose

Five Sigma

Four Sigma

HDL

Six Sigma

Six Sigma

LDL

Six Sigma

Six Sigma

LDH

-

Six Sigma

Lactate

-

Six Sigma

Magnesium

-

Six Sigma

Potassium

Six Sigma

Six Sigma

Total Protein

-

Six Sigma

Triglycerides

Six Sigma

Six Sigma

Urea Nitrogen

Five Sigma

Four Sigma

Uric Acid

Six Sigma

Six Sigma

 The assays and instruments which were verified

 Six Sigma: World Class Quality. Possible to use 3s to 3.5s limits and a minimum of controls

Five Sigma: Excellent Quality. Possible to use 3s limits or a 1:3s/2:2s/R:4s Westgard Rule

Four Sigma: Good Quality. Possible to use 1:3s/2:2s/R:4s/4:1s Westgard Rule

 Visual Summary of Performance of Hoan My Sai Gon Hospital Lab

 2020 hoan may sai gon cc 1 medx

 2020 hoan may sai gon cc 2 medx

 

Congratulations to the entire laboratory staff of the Hoan My Sai Gon hospital.