Tools, Technologies and Training for Healthcare Laboratories

Hoan My Da Nang Hospital Laboratory Sigma Verification of Performance

The laboratory of the Hoan My Da Nang hospital laboratory in Vietnam has joined the Sigma VP program.

Sigma Verification of the Laboratory of the Hoan My Sai Gon Hospital

Duration of Verification: February 1st, 2019 through February 28th, 2020

"The first hospital in Da Nang was granted an international standard clinical test (ISO 15189: 2012)

"Founded in 2001, the Laboratory has become one of the most prominent and prominent features of Hoan My Da Nang Hospital with many departments such as: Hematology - Blood coagulation - Blood transfusion / Blood bank - Biochemistry - Immunology - Molecular Biology - Microbiology - Anatomy of diseases.

"Hoan My Danang Hospital has constantly improved its system of facilities (facilities, facilities, equipment to ensure accurate results) and investment in people ( doctor degree, management capacity and expertise). In addition, the hospital also constantly performs internal, external control, equipment calibration ... in order to improve its operation as well as ensure that it meets the testing quality standards."

The assays and instruments which were verified

Chemistry Assay

Verified on Architect

C804543 ci8200

Verified on Architect

C401736 ci4100

Albumin

Six Sigma

Four Sigma

Bilirubin, Direct

Four Sigma

Four Sigma

Calcium

Four Sigma

-

C-reactive Protein

Six Sigma

Six Sigma

Chloride

Four Sigma

Four Sigma

Creatinine

Six Sigma

Five Sigma

GGT

Six Sigma

Six Sigma

Glucose

Four Sigma

Five Sigma

HDL

Six Sigma

Six Sigma

Iron

Four Sigma

-

PreAlbumin

Five Sigma

-

LDH

Six Sigma

-

LDL

Six Sigma

Six Sigma

Lipase

Six Sigma

-

Lactate

Six Sigma

-

Magnesium

Six Sigma

-

Potassium

Six Sigma

Six Sigma

Triglycerides

Six Sigma

Six Sigma

Urea Nitrogen

Five Sigma

-

Uric Acid

Six Sigma

Six Sigma

 Six Sigma: World Class Quality. Possible to use 3s to 3.5s limits and a minimum of controls

Five Sigma: Excellent Quality. Possible to use 3s limits or a 1:3s/2:2s/R:4s Westgard Rule

Four Sigma: Good Quality. Possible to use 1:3s/2:2s/R:4s/4:1s Westgard Rule

 Visual Summary of Performance of Hoan My Da Nang Hospital Laboratory

 2020 hoan my da nang sigma vp NMEDX1

 2020 SigmaVP Hoan my da nang medx2

Congratulations to the entire laboratory staff of the Hoan My Da Nang hospital.

Let us know what you're interested in!
Please use this form to request more infromation about 

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input