Tools, Technologies and Training for Healthcare Laboratories

Resources - Page 12

QC Selection Grid, Square 7

MULTIRULE CHOICE:

Mqcsg7

You have few problems, but medically important errors are small and difficult to detect, therefore you need to increase the number of rules to four and apply the 41s rule across two runs (R=2) as a rejection rule. The power curves below show the rejection characteristics within one run (R=1) and across two runs (R=2):


Mpf3

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curves. Where the curves and critical-error meet indicates how much error detection each control rule will provide.

QC Selection Grid, Square 8

MULTIRULE CHOICE:

Mqcsg8

You have few problems, but medically important errors are moderate in size, which means you should use at least three rules that can be applied to the two control measurements in a run (N=2) to improve error detection. You can also add a warning rule that works across two runs (R=2). The power curves below show the rejection characteristics within a run (R=1) and across two runs (R=2):


Mpf3

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curves. Where the curves and critical-error meet indicates how much error detection each control rule will provide.

QC Selection Grid, square 9

Multirule Choice:
Mqcsg9

Very cost-effective operation is possible because you have a very stable method and the medically important errors are large and easy to detect. You can actually use a single 13s rule with N=2 for rejection. If you want to look at data over two runs, you can add a 41s warning rule to alert analysts to prospectively check the method before the next run. The power curve below shows the rejection characteristics:


Mpf4

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curve. Where the curve and critical-error meet indicates how much error detection each control rule will provide.