Tools, Technologies and Training for Healthcare Laboratories

MIOT Hospital Sigma metric Verification of Performance

MIOT Hospital becomes the first laboratory in India to achieve the Sigma Verification of Performance.

MIOT International Laboratory Verification of Performance

Verification: December 1st, 2019 through December 31, 2020

A Quick Description of MIOT Laboratory

From their website: "MIOT International offers the best in health care to patients who have come to trust the hospital as one of India’s best.  A 1000-bed facility offering superlative treatment in 63 specialities, MIOT proudly lays claim to a team of world-class professionals who use our state-of-the-art infrastructure and cutting-edge technology to heal and give hope to our patients. Consistently adjudged as a leader in healthcare, MIOT International attracts patients from 130 countries who seek our expertise in moving towards wellness"

Assays Verified at MIOT Laboratory

 Clinical Chemistry assays were verified on the Ortho Clinical Diagnostics instruments VITROS 7600 XT and VITRO 5600.

Immunoassays were verified on Abbott instruments ARCHITECT ci8200, i1000SR, i2000SR

Chemistry Assay

Verified on Vitros 7600XT

Verified on Vitros 5600

Albumin

Six Sigma

Six Sigma

Alk Phos

 

Six Sigma

ALT

 

Five Sigma

AST

Six Sigma

Five Sigma

Bilirubin, Direct

Four Sigma

 

Bilirubin, Total

Six Sigma

Four Sigma

Calcium

Six Sigma

Six Sigma

Chloride

Four Sigma

Six Sigma

Cholesterol

Five Sigma

Five Sigma

CO2

 

Four Sigma

Creatinine

Six Sigma

Six Sigma

GGT

Six Sigma

Six Sigma

Glucose

Six Sigma

Six Sigma

HDL

Six Sigma

Six Sigma

LDH

Five Sigma

Six Sigma

LDL

Six Sigma

Six Sigma

Lipase

Six Sigma

Six Sigma

Magnesium

Six Sigma

Six Sigma

Potassium

Six Sigma

Six Sigma

Phosphorous

Six Sigma

Four Sigma

Total Protein

Six Sigma

---

Triglycerides

Six Sigma

Six Sigma

Uric Acid

Six Sigma

Six Sigma

 

Immunoassays

Verified on ARCHITECTs

Anti-TPO

Six Sigma (i1000SR)

Folate

Six Sigma (ci8200)

CA 125

Six Sigma (i2000SR)

Visual Display of Assay Performance at MIOT Laboratory

 2020 MIOT VP Normalized MEDX chart 1

 

 2020 MIOT VP Normalized MEDX chart 2

 2020 MIOT VP NMEDX3

 

 WSVP Certificate MIOT 2019 2020

Congratulations MIOT Laboratory!